Power on and loud buzzing noise heard. Pull side cover and found noise is from the fill/drain valve. Removed and installed new. Restarted unit. Now is quiet. Unit fills and hats. 25 amp per element and drying element is 1.3 amp. Total amp is 43-44-44 L1, L2, L3. Power off and drain and restart. It fills and heats. Back to steam. Timer times down and sounds ok. In service. ** Travel Charges billed on WO#138004